2.3.1
Voltage generation (DC)
•
Display the
TRANSMISSION CONFIGURATION
dialogue box.
•
Select the
Vdc
transmission function, then the range
suitable for measurement using the function and
navigation keys.
•
Confirm with ENTER.
The following ranges are available:
Range
100 mV
2 V
20 V
Resolution
1 µV
10 µV
100 µV
Output impedance
< 1 Ω
< 1 Ω
< 1 Ω
Output load
1 k
Ω
2
kΩ
4
kΩ
The DC source voltage to be generated is connected between terminals V and COM.
2.3.2
Current generation (DC)
•
Display the
TRANSMISSION CONFIGURATION
dialogue box.
•
Select the
Idc
measurement function, then the range using the function and navigation
keys.
•
Confirm with ENTER.
Depending on the range selected, several transmission modes are available:
Range
24 mA 4-20 mA
0-20 mA
Resolution
1 µA
1 µA
1 µA
Loop power supply
no
possible
possible
Set to scale
no
linear or square law
linear or square law
Connection is made between the mA and COM terminals.
If the loop power supply is on, the MC 75.2 simulates a
passive transmitter supplied with 24 V internally.