Troubleshooting
Instruction Manual LabScat 2
12112E/2
47
The following warning messages can be displayed:
WARNING
DESCRIPTION
POSSIBLE CAUSES
HIGH STANDARD
DEVIATION
The standard deviation is
greater than 50 EBC and
greater than 15% of the
mean value.
Sample is unstable. This can
stem from particles or gas
emissions.
ADJUSTFAULT
The adjustment of the instru-
ment could not be performed.
The instrument is soiled. Sec-
tion 9.2.
The set point for the adjust-
ment does not match the value
of the sample.
WATER SHORTAGE
A float switch is in the water
bath. It reports when the wa-
ter level in the water bath is
too low.
Regulator valve is closed or wa-
ter connection is not correct.
Can be briefly displayed if a
large bottle has been removed
from the water bath.
Float switch is stuck. Section
9.2.
Defect in the electronic system.
Service technician
WATER QUALITY
The water quality is insuffi-
cient.
The water quality in the water
bath is insufficient.
A sample with a high turbidity
value has been in the water
bath for a long time.
MOTOR
The speed of the motor for
the bottle centering device is
too low
Blocking of the bottle center-
ing device. Section 9.2
Defective drive motor.
Defective drive belt.
HUMIDITY
The relative humidity in the
instrument has risen above
the set limit
The desiccant is saturated.
Gaskets on the electronic com-
ponent defective.
Instrument was open too long.
OVER TEMP
The temperature in the in-
strument has exceeded 65 °C.
The medium or ambient tem-
perature is too high and defec-
tive or no cooling.
Table 3: Possible warning messages