Instruction Manual LabScat 2
Troubleshooting
48
12112E/2
10.3
Error messages and effect on operation
FAULT
If an error occurs during operation, it has the following
effects:
The measurement is canceled.
The measuring values go to 0.
When the
Fault
message appears,
the color of the
status display changes to
red
and the text de-
scribes the fault in question.
If the cause of the fault has been remedied, it is
automatically deleted.
Example:
TOO MUCH EXTERNAL
LIGHT
The following error messages can be displayed:
ERROR MESSAGE
DESCRIPTION
POSSIBLE CAUSES
V ANALOG
One of the internal analogue
voltages is outside the permit-
ted range.
Defect in the electronic system.
Service technician
MEASUR.FAULT
Measurement value acquisi-
tion is faulty.
Sample very unstable.
Large air bubbles in the water
bath.
External light influence too
strong. Water bath cover is
open.
Defect in the electronic system.
Service technician
LED1 FAILURE
The detector for the LED mon-
itoring is not receiving any
light.
Defective light source.
Service technician
TOO MUCH
EXTERNAL LIGHT
The detectors receive too
much light.
Water bath cover open.
ABSORPTION TOO
HIGH
The sample absorbs the light
too strongly. A measurement
is not possible.
The sample and/or the bottle is
too dark.
HUMIDITY
The relative humidity in the
instrument has risen above
50%.
The desiccant is saturated.
The seals on the electronic
component are defective.
The instrument was open for
too long.
Table 4: Possible error messages