Servicing
Instruction Manual LabScat 2
12112E/2
43
9.3
Adjustment with checking unit
The following procedure describes adjustment with the control unit:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Clean water bath.
Section 9.2
2.
Ensure that the water supply to the water bath
corresponds to the specifications.
Section 2.4
3.
Fill water bath.
Section 4.3
4.
Inspect the checking unit of the instrument
and ensure that it is clean and undamaged.
5.
Check whether the units specified on the
checking unit match the nominal values in the
instrument. This can be done as follows:
a:
Press the
Menu
button.
b:
Enter access code.
c:
Press the
Recalibr.
button and then
C1 90°
and check whether the saved nominal value of
90° matches the specification on the checking
unit.
d:
Press the
ESC
button.
e:
Press the
C2 25°
button and check whether
the saved 25° nominal value matches the spec-
ification on the checking unit.
f:
Press the
ESC
button.
6.
Put the checking unit into the water bath and
close the cover.
7.
Prepare adjustment.
Press the
C1 90°
button.