Instruction Manual DualScat Ex
38
10118E/5
6
Troubleshooting
6.1
Pinpointing causes of trouble
Work through the following table step by step to narrow down the possible
causes of a malfunction. If the listed corrective steps fail to produce the de-
sired results, please consult Customer Service.
Section 6.4
Perceived malfunction
Corrective steps
No display
-
Make sure the supply voltage is actually ap-
plied to the control unit.
Section 3.2
-
Check fuses in the control unit.
Reference Handbook
Fault message displayed
-
Analyze the fault message.
Section 6.2
Reading is very suspect
-
Check whether the medium in the product pipe
still meets the normal duty conditions.
Section 1.5
-
Carry out a sensor check.
Section 6.3
-
Check whether the photometer is properly in-
stalled.
Section 3.1
-
Check whether the service work has been car-
ried out according to the Service Schedule.
Section 5.1
-
Check the dry chamber in the Ex enclosure.
Section 5.2
-
Check the dry chamber in the sensor head.
Section 5.3
-
Clean the sensor head.
Section 5.4
-
Calibrate the photometer.
Section 5.6
No measured value output
signal (0 mA)
-
This status signals that a fault has occurred in
the photometer.
-
Analyze the fault displayed fault message.
Section 6.2
Содержание DualScat Ex
Страница 2: ... SIGRIST PHOTOMETER AG Subject to change without notice 9 2016 ...
Страница 6: ...Instruction Manual DualScat Ex iv 10118E 5 ...
Страница 53: ...Instruction Manual DualScat Ex 10118E 5 47 ...
Страница 54: ...Instruction Manual DualScat Ex 48 10118E 5 11 Appendix ...
Страница 55: ...Instruction Manual DualScat Ex 10118E 5 49 ...