Instruction Manual DualScat Ex
10118E/5
43
7
Taking Out of Service/Storage
The objective of this procedure is proper preparation of the photometer for
storage and retention of its normal condition during the storage period.
You must interrupt the service voltage (explosion hazard) before opening the
SIREL Ex housing.
Action
1.
Disconnect the power supply to the control unit and remove all electrical
connections from the control unit.
2.
Shut off the flow in the product pipe and drain it.
3.
Remove the photometer from the product pipe and clean the sensor
head.
Section 5.4
4.
Thoroughly clean all surfaces that have come into contact with the
medium. Now no toxic, corrosive or loose deposits should remain inside.
5.
Check the photometer's dry chamber and if necessary replace the desic-
cant.
Section 5.2
6.
Demount the control unit.
7.
Make sure all covers and hoods for the photometer and control unit, and
all closures on them, are properly locked.
No particular conditions are required for storage of the equipment, but do ob-
serve the following points:
Because the photometer and control unit contain electronic components,
the conditions should meet the normal requirements for such components.
In particular, the storage temperature should be within the -20 .. +50°C
range.
All components that have come into contact with the medium during oper-
ation must be dry and clean for protracted storage or shipment.
The photometer, control unit and accessories must be protected against
weathering, condensing moisture and aggressive gases while in storage.
Taking out of
service
Содержание DualScat Ex
Страница 2: ... SIGRIST PHOTOMETER AG Subject to change without notice 9 2016 ...
Страница 6: ...Instruction Manual DualScat Ex iv 10118E 5 ...
Страница 53: ...Instruction Manual DualScat Ex 10118E 5 47 ...
Страница 54: ...Instruction Manual DualScat Ex 48 10118E 5 11 Appendix ...
Страница 55: ...Instruction Manual DualScat Ex 10118E 5 49 ...