Instruction Manual DualScat Ex
34
10118E/5
Action
5.
Insert the photometer into the checking unit
and clamp it firmly in place.
6.
Turn the entire assembly to the horizontal
position and place the filling funnel on top
(
Figure 16).
7.
Switch on the power supply to the control
unit and let the instrument warm up for at
least 3 minutes.
8.
Fill in distilled water carefully through the
funnel (A) until the level indicator is about
half full of water (B).
Make sure no bubbles are created when the
water flows into the checking unit, so that
no troublesome bubbles occur inside it.
9.
Carry out the recalibration on the control unit
as directed in Section 5.6.3.
The nominal value(s) are
printed on the checking
unit.
10.
Drain the checking unit, remove the photom-
eter and replace it in your product pipe.
Section 3.1
11.
Restart the instrument in normal operation.
12.
Clean and dry the checking unit, inside and
out, with a soft, lint-free rag. If severely
fouled, use an acidic, nonabrasive cleaning
agent.
Do not use any other cleaning methods or
cleaning agents!
A
B