Instruction Manual DualScat Ex
10118E/5
37
Action
Display (Example)
Remarks
8.
> Val.Adjust 90<
34.5 NTU
For single-angle instru-
ment, proceed to Point 12.
9.
> Val.Adjust 25<
0.00 NTU
10.
Enter nominal value:
/
change number
/
change place
Val.Adjust 25
> ▓.00 NTU <
Now enter the nominal
value of the checking unit
for the relevant measuring
angle or the precise value
of the calibration suspen-
sion in the NTU unit.
11.
> Val.Adjust 25<
3.50 NTU
12.
-Meas90 -Meas25
34.5 3.50
Display of the detected
values without corrections.
If they are in line with the
previously entered nominal
values, no recalibration is
necessary; proceed to
Point 18.
13.
> Adju. <
Retain
14.
Adju.
> Retain <
15.
Adju.
> Adapt <
16.
Adju.
running ...
Compute the calibration
factors.
17.
Wait until the
DualScat Ex has
computed the
calibration factors
and the display
changes
-Meas90 -Meas25
34.5 3.50
Display of the "rough" ref-
erence values without de-
tected value corrections.
If "outside tolerance" ap-
pears on the display, the
nominal values and detect-
ed values differ too widely.
If this happens, check
whether the entered nomi-
nal values agree with the
values of the calibration
suspension or the checking
unit.
18.
+
(simultaneous)
90° 34.5 NTU 1
25° 3.50 NTU 1
Instrument in normal oper-
ation.
The entered nominal values are stored internally and remain un-changed until
the next recalibration.
Содержание DualScat Ex
Страница 2: ... SIGRIST PHOTOMETER AG Subject to change without notice 9 2016 ...
Страница 6: ...Instruction Manual DualScat Ex iv 10118E 5 ...
Страница 53: ...Instruction Manual DualScat Ex 10118E 5 47 ...
Страница 54: ...Instruction Manual DualScat Ex 48 10118E 5 11 Appendix ...
Страница 55: ...Instruction Manual DualScat Ex 10118E 5 49 ...