18. Remove the optics unit from the AquaScat SG (A)
and place in the recess (Dockingstation).
Section 9.4.1
19. Check the water jet.
Section 9.4.2
20. Check that the light trap and shutter are clean.
Section 9.4.2
21. Place the optics unit back on the measuring cell
unit of the AquaScat SG (A) and lock with the
four mounting clips. Pay attention to the guide
pins (see figure).
22. Check the sample flow in the OilGuard SG (A).
Section 9.4.2
23. Remove the optics unit from the OilGuard SG (A)
and place in the recess (Dockingstation).
Section 9.4.1
24. Check the water jet.
Section 9.4.2
25. Check that the light trap and shutter are clean.
Section 9.4.2
26. Place the optics unit back on the measuring cell
unit of the OilGuard SG (A) and lock with the four
mounting clips. Pay attention to the guide pins
(see figure).
27. Check the ScrubberGuard for leaks according to
the checklist in the servicing schedule.
Section 9.2.1
28. Check the sample flow in the system.
Section 9.2.2
29. The ScrubberGuard is now ready for operation.