Instruction Manual ScrubberGuard
Servicing
86
14830E/1
WORKSTEP
ADDITIONAL INFO / IMAGES
8.
Position the mounting plate with the newly
mounted valves and fasten in place with the
four screws (arrows).
9.
Fasten the pipe on both valves at the valve
inlet and valve outlet. In doing so, ensure that
the pipes are not under tension.
10. Fasten both valves onto the mounting plate
from below using the four fastening screws.
11. Plug in the connector on both valves and fas-
ten in place with the screw (arrow).
12. Re-establish the sample feed to the system.
13. Reattach the casing to the system.
14. The system is now ready for operation again.