The following fault messages can be displayed:
FAULT MESSAGE
DESCRIPTION
POSSIBLE CAUSES
V ANALOG
One of the internal analog
voltages is outside the permit-
ted range.
Defect in the electronic system.
Service technician
MEASUR.FAULT
Measuring value acquisition is
faulty.
Bubbles in the water.
External light in the vicinity of
the measuring point (e.g.
transparent hoses).
Defect in the electronic system.
Service technician
AN.MEAS.FAULT
The measuring value acquisi-
tion of the analog channels is
disturbed.
Defect in the electronic system.
Service technician
LIGHTSOURCE 1
The detector for monitoring
the light source receives no
light from the corresponding
light source.
Defective light source.
Service technician
SERIAL 1 .. 8
The photometer cannot estab-
lish a connection to the Hamil-
ton sensor (slave number 1 ..
8).
Interrupted connection to sen-
sor 1 .. 8.
Defect in the electronic system.
Service technician
WATER SHORTAGE
The flow rate is too low.
Too little water in the sample
line.
Feed pump is defective.
Inlet valve is blocked.
MEASURING
Measuring problem with sen-
sor.
Temperature or measuring val-
ues are unstable or outside the
permitted range.
INTERFACE
Connection problem with
Hamilton sensor.
mA value outside range.
ECS (electrical connection of
this sensor) is outside the
range.
HARDWARE
Hardware problem with Ham-
ilton sensor.
Supply voltage outside range.
CALIBRATION
Calibration problem with
Hamilton sensor.
Calibration recommended.
Last calibration not successful.