9.2.2
Checking the flow rate on the system
This section relates to the sample volume on the inlet and outlet off the tank.
The inspection must be made during regular measuring operation.
When checking the flow rate of the instruments, please consult the Section 9.4.2.
9.2.2.1
Flow rate too low:
If the flow rate is too low, the
WATER SHORTAGE
fault message appears.
9.2.2.2
Flow rate too high:
An excessive flow rate can be determined as follows:
WORKSTEP
ADDITIONAL INFO / IMAGES
1.
Drain the water tank by pressing the
Drain
tank
button.
The tank cover may be removed for better as-
sessment (arrow).
The pump may only be oper-
ated dry for a maximum of 60
seconds. Danger of overheating!
2.
Pressing the
Measurement ON/OFF
button
starts filling of the tank.
If the filling time is less than 20 seconds, then
the flow rate is too high.