Using the Centrifuge – Spincontrol Professional
46
Version 12/2008, Rev. 1.11 of 27/01/2012
Translation of the original operating manual
Runtime
The set runtime is displayed in the upper section of this field, with the
remaining runtime shown below. The runtime is defined as the period
from the start of the centrifuge to the beginning of the deceleration
phase. The maximum value is 99 h 59 min 59 sec.
In the “Setup” menu, one can specify that the counting of the runtime
is not to be started until the set speed is reached (see 6.4.3.8 "Setup
Menu"). In this case, the symbol "
!
" appears in the speed field.
Continuous Run
During the continuous run, the runtime of the centrifuge is unlimited
and must be stopped manually. The centrifuge accelerates during the
continuous run until the set speed is reached.
Select the field “Runtime” and confirm the selection. The display
flashes when it is activated
Select the time 0:00:00 and confirm the selection. "Infinite" will be
displayed. After the start of the centrifuge, the elapsed time will be
displayed.
Deactivate the continuous run by pressing the stop key or by
entering a specific run time.
Short Run
Keep the start key pressed during the short run.
During the short run, the centrifuge accelerates with the maximum
acceleration curve until the maximum speed of the rotor is reached.
When the start key is released, the centrifuge decelerates at
maximum brake to a standstill.
Rotor: Rotor Selection List/ Automatical Rotor Identification
This field shows the rotor that is currently being used.
Select the field "Rotor" and confirm the selection. A list with all of
the possible rotors without buckets is displayed.
Select the desired rotor. All of the possible rotor/bucket
combinations and additional information for each combination will
be displayed.
Select a rotor/bucket combination and confirm the selection.
Press the function knob in order to accept the data.
Fig. 6.30: Rotor Selection List