![Sigma 3-30K Скачать руководство пользователя страница 30](http://html1.mh-extra.com/html/sigma/3-30k/3-30k_operating-manual_1267902030.webp)
3-30K
Using the Centrifuge – Spincontrol Comfort
30
Version 12/2008, Rev. 1.11 of 27/01/2012
Translation of the original operating manual
The set value is indicated in hours, minutes, and seconds. The actual
value has the same units as the set value and is displayed in
hours:minutes, or in minutes:seconds if the value is below 10
minutes. Values greater than 59 will be automatically converted to the
next higher unit.
Continuous Run
During the continuous run, the runtime of the centrifuge is unlimited
and must be stopped manually. The centrifuge accelerates during the
continuous run until the set speed is reached.
To start the continuous run of the centrifuge :
Starting with the time setting 0:10 (see page 29), turn the knob
counter-clockwise or
Starting with the time setting 09:59, turn the knob clockwise to the
next setting. "HOLD” will be displayed. After the start of the
centrifuge, the elapsed time will be displayed.
Fig. 6.10: Display of a continuous run
Deactivate the continuous run by pressing the stop key or by
entering a specific runtime.
Short Run
Keep the start key pressed during the short run.
During the short run, the centrifuge accelerates with the maximum
acceleration curve until the maximum speed of the rotor is reached.
When the start key is released, the centrifuge decelerates at
maximum brake to a standstill.
Temperature
The set value is in the upper section of the area, underneath the
actual temperature is displayed. Temperatures between -20 °C and
+40 °C can be preselected.
In the centrifuges without an active heater, the temperatures above
room temperature depend on the air friction of the turning rotor.
Fig. 6.11: Setting the temperature