3-30K
Using the Centrifuge
Version 12/2008, Rev. 1.11 of 27/01/2012
25
Translation of the original operating manual
6.2.2.2
Installation of Angle Rotors with a Hermetically Sealed Lid
Screw the rotor cover onto the rotor and tighten it.
Lower the rotor with the cover onto the motor shaft .
Insert the rotor tie-down screw into the motor shaft. Tighten the
rotor tie-down screw with 10 Nm using the supplied rotor wrench
so that the spring washer assembly is compressed tightly.
The rotor can also be used without a cover.
Slightly grease the rotor and lid seals after cleaning.
The rotors can be installed or removed with a closed lid after
loosening the rotor tie-down screw.
Follow the safety instructions and hazard warnings in chapter 3!
1 Rotor tie down
screw
2 Lid
3 Lid seals
Fig. 6.1: Angel rotor with a hermetically sealed lid
6.2.2.3 Installation of Accessories
Only use inserts that are suitable for the rotor (see 11.1 "Suitable
Accessories").
All four buckets of the swing-out rotor need to be installed when
spinning.
Always load the opposite inserts/buckets of the rotors with the
same accessories and fill to avoid imbalance.
Centrifugation with different tube sizes:
Working with different tube sizes is possible. In this case, however, it
is very important that opposite inserts are identical (see Fig. 6.2).
Centrifugation with low capacity: The tubes must be installed
symmetrically so that the buckets and their inserts are loaded evenly
(see Fig. 6.3).