6.3.3
Program mode
A program contains all the data that are required for a centrifuge run.
Certain sedimentation results can be repeated under identical conditions.
Programs can be saved, loaded, executed, and edited, when the centrifuge
is at a standstill.
A maximum of 50 programs can be stored under the numbers 1 - 50.
"--" means that the values that are currently set are not a stored program.
The programs can be protected against unauthorised use, modification, or
deletion with the aid of an input lock (see chapter 6.3.2.13 - "Input lock").
Pos : 148 /010 Uni vers almodul e/ Leerz eile @ 0\ mod_1202116244500_0.doc x @ 114 @ @ 1
Pos : 149 /100 Sig ma/100 BA Z entrif ugen Sig ma (St andardmodul e)/060 Betrieb_Spi nc ontr ol L+S/ Spinc ontrol L (Z ent6) /060- 0030-0030- 0011 Programm speic her n - ohne T emp @ 27\ mod_1405319311134_68. doc x @ 193039 @ 4 @ 1
6.3.3.1
Saving a program
NOTE
This function is only available if the centrifuge is at a standstill.
• Enter the parameters that are to be included in the program.
• Select the menu item "save Prog" and confirm the selection. The
indication "save" flashes in the activated status.
• Select a random storage location from the program selection list. Free
storage locations are indicated by a flashing display. Any storage
locations that are already occupied will be overwritten during the saving
process.
• Save the program in the desired location.
The program is now saved.
NOTE
The functions "Auto Open", "Buzzer", "ProgLock", and "Cycles" cannot be
entered as part of a program.
Fig. 24: Saving a program