Laboratory Centrifuge Sigma 3-16L
Laboratory Centrifuge Sigma 3-16L IVD
4 Storage and transport
Version 03/2014, Rev. 2.11 of 11/08/2017
• sb
25 / 79
Translation of the original operating manual, part no. 0700902
Pos : 84 / 100 Sigma/ 110 BA Zentrif ugen Sigma (Pr ojekte) /3- 16L/ 040 Lagerung und Tr ans port/ 040-0040 Verpac kung 3- 16L @ 29\ mod_1405321600134_68.doc x @ 197197 @ 2 @ 1
4.4
Packaging
The centrifuge is packaged in a slip-lid box.
• Take off the lid.
• Remove the box containing the accessories and the packaging material.
• Remove the slip-lid box.
• Lift the centrifuge upwards with a lifting device or with a sufficient
number of people to lift it safely. When lifting the centrifuge, always
reach under the centrifuge from the side.
CAUTION
The centrifuge weighs approx. 48 kg!
• Retain the packaging for any possible future transport of the centrifuge.
Pos : 85 / 010 U ni vers al module/ Leerz eile @ 0\ mod_1202116244500_0. doc x @ 114 @ @ 1
Pos : 86 / 100 Sigma/ 110 BA Zentrif ugen Sigma (Pr ojekte) /3- 16L/ 040 Lagerung und Tr ans port/ 040-0050 Trans port sicherung 3- 16L - Übersc hrift @ 29\ mod_1405321600880_68.doc x @ 197211 @ 2 @ 1
4.5
Transport safety device
Pos : 87 / 100 Sigma/ 110 BA Zentrif ugen Sigma (Pr ojekte) /3- 16L/ 040 Lagerung und Tr ans port/ 040-0051 Trans port sicherung 3- 16L - Text @ 36\ mod_1434605535989_68.doc x @ 263887 @ @ 1
The transport safety device consists of a foamed plastic piece in the rotor
chamber.
CAUTION
The transport safety device must be removed prior to start-up because the
screws lock the motor bearings!
Removal
• Open the lid by pressing the lid key. If the centrifuge is not connected to
the power supply, use the emergency release of the lid (see chapter
7.1.1 - "Emergency lid release").
• Remove the foamed plastic piece (see figure, item 1) from the rotor
chamber, by lifting it carefully on one side.
Fig. 5: Removal of the transport safety device
• Retain the transport safety device for the possibility of the return of the
centrifuge.