Laboratory Centrifuge Sigma 3-16L
Laboratory Centrifuge Sigma 3-16L IVD
6 Using the centrifuge
30 / 79
Version 03/2014, Rev. 2.11 of 11/08/2017
• sb
Translation of the original operating manual, part no. 0700902
Centrifugation with low capacity
• Install the tubes axial symmetrically so that the buckets and their inserts
are loaded evenly.
• It is not permissible to load angle rotors on only one axis.
correct
incorrect
Fig. 8: Permissible and impermissible loading of an angle rotor and a swing-out rotor
(example illustrations)
Pos : 112 /010 Uni vers almodul e/ Leerz eile @ 0\ mod_1202116244500_0.doc x @ 114 @ @ 1
Pos : 113 /100 Sig ma/100 BA Z entrif ugen Sig ma (St andardmodul e)/060 Betrieb_Spi nc ontr ol L+S/ 060- 0020-0020-0040 Adapt er - nur L+S @ 27\ mod_1405319210191_68. doc x @ 192563 @ 4 @ 1
6.2.2.4
Adapters
In order to ensure easy handling, even if vessels of various sizes are used,
carrier systems were developed.
• Load the opposite adapters with the same number of vessels and with
the same weights in order to avoid imbalance.
• If all of the compartments of a carrier are not used, the buckets must be
loaded evenly. Loading the edges of a bucket only is not permissible.
Pos : 114 /010 Uni vers almodul e/ Leerz eile @ 0\ mod_1202116244500_0.doc x @ 114 @ @ 1
Pos : 115 /100 Sig ma/100 BA Z entrif ugen Sig ma (St andardmodul e)/060 Betrieb_Spi nc ontr ol L+S/ 060- 0020-0020-0050 Gefäß e ( auß er 4-5L) @ 27\ mod_1405319211622_68.doc x @ 192577 @ 4 @ 1
6.2.2.5
Tubes
• Load the tubes outside of the centrifuge. Liquids in the buckets or
multiple carriers cause corrosion .
• Fill the tubes carefully and arrange them according to their weight.
Imbalances result in the excessive wear of the bearings.
• In high-speed angle rotors, the vessels must be filled up to their useful
volume (= the volume stated for the vessel). If the vessels are only
partially filled, they will deform. This may result in leaks at the seals that
may become loose.
• When using glass tubes, the maximum value of 4,000 x g must not be
exceeded (except special high-strength glass tubes; please refer to the
information provided by the manufacturer).
• Follow the safety instructions and hazard warnings (see chapter 3 -
Pos : 116 /010 Uni vers almodul e/ Seit enwechs el @ 0\ mod_1202116244312_0.doc x @ 105 @ @ 1