Operating Manual SIGMA 2-16PK
page 39 of 77
05/2007
5.3.3
Relative Centrifugal Force (RCF)
The relative centrifugal force (RCF) is the acceleration that the sample is subject
to.
The RCF value is determined by the rotor geometry and speed. The RCF and
speed values, therefore, depend on each other. When the maximum permissible
RCF value of the rotor is reached, the "max" display lights up.
To preselect the RCF value:
•
Turn the left rotary knob until “set” appears in front of the parameter “rcf” on
the display (Fig. 5.14).
•
Select the option by pressing or turning the right rotary knob. "Set" will now
flash.
•
Turn the right rotary knob until the desired value is displayed.
•
Press the right rotary knob to confirm the selected value. If this is not done,
the value will be automatically reset to the last setting.
It is also possible to change the RCF value during a centrifugation run.
5.3.3.1 Changing the RCF Increments
In order to change the RCF value in steps of 1 x g (instead of in steps of 10 x g):
•
Keep the stop key pressed while setting the desired RCF value with the right
rotary knob.
Fig. 5.14:
Setting the relative centrifugal force
Содержание 2-16PK
Страница 1: ...Refrigerated Centrifuge 2 16PK Operating Manual From Serial No 120951 Please retain for later use...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 12: ......
Страница 14: ......
Страница 65: ...Operating Manual SIGMA 2 16PK page 65 of 77 05 2007 Fig 11 1 Speed gravitational field diagram...
Страница 68: ......
Страница 70: ......
Страница 74: ......