5.2.3.1 Installation of Angle Rotors with a Hermetically Sealed Lid
•
Screw the rotor cover onto the rotor and tighten it.
•
Lower the rotor with the cover onto the motor shaft.
•
Insert the rotor tie-down screw into the motor shaft. Tighten the rotor tie-down
screw at 5 Nm using the supplied rotor wrench so that the spring washer
assembly is compressed tightly.
•
The rotor can also be used without a cover.
•
Slightly grease the rotor and lid seals after cleaning.
•
The rotors can be installed or removed with a closed lid after loosening the
rotor tie-down screw.
)
Attention!
•
Please follow the safety instructions and hazard warnings in chapter 2,
page 16 ff!
Fig. 5.5:
Angle rotor with a hermetically sealed lid
1 Rotor tie-down screw
2 Lid
3 Lid
seals