5.2.4
Installation of Accessories
•
Only use vessels that are suitable for the rotor (see chapter 10 "Suitable
Accessories", page 56).
•
Load all of the compartments of the swing-out rotors.
•
Always load the opposite compartments of the rotors with the same
accessories and filling to avoid imbalance.
•
Centrifugation with low capacity: For the
example, a swing-out rotor
(16 x 15 ml) that is to be equipped with
eight tubes only was used. The tubes
must be installed symmetrically so that the
buckets and their suspensions are loaded
evenly.
•
Centrifugation with different tube sizes:
Working with different tube sizes is
possible. In this case, however, it is very
important that opposite inserts are
identical. The example shows a swing-out
rotor with buckets and 2 x the insert 100
ml and 2 x the insert 4 x 15 ml with the
corresponding glass or plastic tubes.
Another example is the angle rotor 24 x 2.2 ml: The tubes must be installed
symmetrically so that the rotor is loaded evenly.
Fig. 5.6: Permissible loading of
the swing-out rotor
Fig. 5.8:
Permissible loading of the angle rotor
Fig. 5.7:
Permissible loading of the swing-
out rotor with different tube sizes
Содержание 2-16PK
Страница 1: ...Refrigerated Centrifuge 2 16PK Operating Manual From Serial No 120951 Please retain for later use...
Страница 2: ......
Страница 4: ......
Страница 6: ......
Страница 12: ......
Страница 14: ......
Страница 65: ...Operating Manual SIGMA 2 16PK page 65 of 77 05 2007 Fig 11 1 Speed gravitational field diagram...
Страница 68: ......
Страница 70: ......
Страница 74: ......