Blood bank centrifuge Sigma 8KBS
5 Set-up and connection
Version 05/2021, Rev. 1.2 of 28/04/2022
• sb
31 / 125
Translation of the original operating manual, part no. 0703502
→
5
Set-up and connection
The centrifuge must be set up and connected by authorised specialised
personnel. All the relevant information is documented in a separate
transport and installation manual.
NOTE
• Comply with the information given in the separate transport and
installation manual of the centrifuge!
• Ensure that any tasks concerning this field are only performed by
authorised specialised personnel!
→
5.1
Condensate drain
The condensate drain is used to drain the condensate off that forms in the
rotor chamber during the centrifugation. It consists of a hose with a plastic
valve. It consists of a hose with a plastic valve and runs from the rotor
chamber to the outlet on the left side in the front door of the centrifuge (see
chapter 2.1.1 - "Functional and operating elements").
WARNING
• Do not open the condensate drain unless the rotor is at a complete
standstill.
Draining the condensate off
• Switch off the centrifuge at the mains switch and unplug the mains
power plug.
• Connect the supplied hose connector and drain the condensate.
• Disconnect the hose connector by pressing the unlocking button.