Blood bank centrifuge Sigma 8KBS
6 Using the centrifuge
Version 05/2021, Rev. 1.2 of 28/04/2022
• sb
37 / 125
Translation of the original operating manual, part no. 0703502
NOTE
Pay attention to the marking of the centrifuge (see the illustration below)!
Safety indications on the centrifuge must be kept readable at all times. If
necessary, they must be replaced.
rotor:
bucket:
Fig. 16: Safety indication on the centrifuge: Loading of a swing-out rotor
→
6.2.2.3
Adapters
In order to ensure easy handling, even if vessels of various sizes are used,
carrier systems were developed.
• Load the opposite adapters with the same number of vessels and with
the same weights in order to avoid imbalance.
• If all of the compartments of a carrier are not used, the buckets must be
loaded evenly. Loading the edges of a bucket only is not permissible.
→
6.2.2.4
Vessels and blood bag systems
NOTE
The information and instructions provided by the manufacturers of bags,
bottles and tubes for use in the centrifuge, and especially the information
concerning the maximum permissible RCF value and temperature, must
be observed.
Vessels
• Load the tubes outside of the centrifuge. Liquids in the buckets or
multiple carriers cause corrosion.
• Fill the vessels carefully and arrange them according to their weight.
Imbalances result in the excessive wear of the bearings.
• Always fill the tubes up to their useful volume (= the volume that is
stated for the tube).
• After the centrifugation, remove the vessels carefully in order to prevent
the samples from mixing.
• Follow the safety instructions and hazard warnings (see chapter 3 -