Blood bank centrifuge Sigma 8KBS
6 Using the centrifuge
Version 05/2021, Rev. 1.2 of 28/04/2022
• sb
59 / 125
Translation of the original operating manual, part no. 0703502
Deleting the stored barcodes: "deletebarcodes"
This command is used to delete all of the stored barcodes.
The "Barcode" menu will be updated.
NOTE
This command will not be accepted unless the centrifuge is at a standstill.
Reading out the data of the last centrifugation run: "getlastrun"
The response to this command includes information concerning the
parameters and status of the last centrifugation run (for the exact format of
the data see chapter 11.6 - "Serial Control Interface Specification").
If barcodes have been stored, they will also be output as follows:
• Syntax data structure 001:
"Barcode;x;abc\r\n" (x = serial number starting at 1, abc = barcode)
• Syntax data structure 020:
"Barcode;Staff Member ID;abc\r\n" (abc = barcode)
The enquiry can be performed at any time (before, during, and after the
centrifugation run). Depending on the time of the enquiry, some of the
results may still be unknown.
→
6.3.2.11 Help menu
The help function is symbolised by the " " symbol on the menu bar. It
provides a short description of the control elements of the selected option.
Activating and deactivating the help function
• Select the question mark on the menu bar and press the function knob.
• Quit the help function by selecting the question mark and by pressing
the function knob again.
Parameters can still be changed when the help function is activated.