42
User Manual
Refreshed and displayed trace data.
View
The trace data will not be refreshed, and the current latest trace will be displayed in a fixed frame.
Blank
Trace data is no longer refreshed or displayed
Background
Refreshed trace data but no display.
Command format
:TRACe[1]|2|3|4|5|6:DISPlay[:STATe]
:TRACe[1]|2|3|4|5|6:DISPlay[:STATe]?
Instructions
Sets/gets the display status of the trace
Parameter Type
Enumeration
:
ACTI|VIEW|BLAN| BACK
Parameter Range
ACTIve
:
Trace is in normal mode. Update data
VIEW
:
Stops updating trace to display current trace data
BLANk
:
Clear trace data
BACKground: Set as background
Return
Enumeration
:
ACTI|VIEW|BLAN| BACK
Example
:TRACe2:DISPlay BLANK
:TRACe2:DISPlay?
3.6.4
Detect
The analyzer displays the sweep signal on the screen in the form of a trace. For each trace point,
the analyzer always captures all the data within a specific time interval and processes (Peak,
Average, etc.) the data using the detector currently selected, then it displays the processed data (a
single data point) on the screen.
Select an appropriate detector type according to the actual application in order to ensure the
accuracy of the measurement.
The available types are
Pos Peak
,
Neg Peak
,
Sample
,
Normal
,
Average
and
Quasi Peak
. The
default is
Pos peak
.
Positive Peak
For each trace point, Positive Peak detector displays the maximum value of data sampled within
the corresponding time interval.
Negative Peak
Содержание SSA5083A
Страница 1: ...User Manual EN_01A SSA5085A SSA5083A Spectrum Analyzer...
Страница 2: ......
Страница 4: ......
Страница 29: ...User Manual XXV 10 2 TROUBLESHOOTING 225...
Страница 30: ......
Страница 97: ...User Manual 67...
Страница 253: ...User Manual 223 Power SYSTem POWer OFF SYSTem RESTart Buzzer DISPlay WINDow BEEP STATe DISPlay WINDow BEEP VOLUme...
Страница 259: ......