134
User Manual
4.8.5.1
Average | Hold Times
Average|Hold Times N, which is the counter when the trace type in Density view is "Average",
"Max Hold" and "Min Hold". In a single measurement (Single), and any valid trace type is
"average", "max hold" or "min hold", the sweep stops when the counter reaches N.
Larger average | hold times can reduce the influence of noise or other random signals, thereby
highlighting stable signal characteristics in the signal.
Command format
[:SENSe]:AVERage:TRACe[1]|2|3|4|5|6:COUNt
[:SENSe]:AVERage:TRACe[1]|2|3|4|5|6:COUNt?
Instructions
Sets/gets the average number of traces
Parameter type
Integer
Parameter Range
1 ~ 999
Return
Integer
Example
:AVERage:TRACe1:COUNt 10
4.8.5.2
Persistence
Sets the time when a frequency / amplitude display point's brightness fades in the persistence
bitmap.
⚫
In finite mode, you can customize the afterglow duration. And the length of time that the
brightness of a point decays from 100% to 0%.
⚫
In infinite mode, the display brightness of each point is 100% without attenuation, but its
probability will change with the measurement time.
Command format
:DISPlay:VIEW:DENSity:PERSistence
:DISPlay:VIEW:DENSity:PERSistence?
Instructions
Set the duration of afterglow
Query the duration of the afterglow
Parameter type
Float
,
unit ks
、
s
、
ms
、
us
Parameter Range
0 s ~ 10 s
Return
Float
,
unit s
Example
:DISP:VIEW:DENS:PERS 5s
Command format
:DISPlay:VIEW:DENSity:PERSistence:INFinite
:DISPlay:VIEW:DENSity:PERSistence:INFinite?
Содержание SSA5083A
Страница 1: ...User Manual EN_01A SSA5085A SSA5083A Spectrum Analyzer...
Страница 2: ......
Страница 4: ......
Страница 29: ...User Manual XXV 10 2 TROUBLESHOOTING 225...
Страница 30: ......
Страница 97: ...User Manual 67...
Страница 253: ...User Manual 223 Power SYSTem POWer OFF SYSTem RESTart Buzzer DISPlay WINDow BEEP STATe DISPlay WINDow BEEP VOLUme...
Страница 259: ......