38
User Manual
Command format
[:SENSe]:SWEep:EGATe:DELay
[:SENSe]:SWEep:EGATe:DELay?
Instructions
Set or query the gate delay
Parameter Type
Float
Parameter Range
Swept
:
8.906us~25s
Zero span
:
1.894us~25s
Return
Float
Example
SWEep:EGATe:DELay 0.005
3.5.11.5
Gate View
When Turn on this view:
⚫
Disable the gate on/off and the gate function.
⚫
To enter the zero span mode, set different Gate View Sweep Time based on different gate
methods.When you close the gate view, the sweep span and sweep time before the gate view
is enabled are restored.
Gate View Sweep Time
Controls the sweep time in the gate view window. When selecting different gate methods, the
instrument will automatically configure different sweep time.
Gate View Start Time
Set the start time on the left of the gate view, that is, set a delay.
Command format
[:SENSe]:SWEep:EGATe:VIEW
[:SENSe]:SWEep:EGATe:VIEW?
Instructions
Set or query the gate view switch
Parameter Type
Boolean
Parameter Range
0|1
Return
0|1
Example
:SWEep:EGATe:VIEW 1
Command format
[:SENSe]:SWEep:EGATe:VIEW:STARt
[:SENSe]:SWEep:EGATe:VIEW:STARt?
Instructions
Set/Query the gate View start time
Parameter Type
Float