2.

 

Select the mounting method depending on the pipe diameter. 

V Method: Pipes 50mm to 400mm (2 inches to 16 inches); signal bounces off pipe wall 
 

 

Z Method Pipes:  400mm to 1200mm (16 inches to 48 inches); signal is directly transmitted

 

 

3.

 

Turn on the meter.   

 

4.

 

Enter the pipe setup parameters.  
Note:  

a)

 

The distance (spacing) between the upstream and downstream transducers is calculated using the pipe 
parameters you enter. 

b)

 

Metrics or English units are selected in Menu 30. 

  

Example: Let us assume you have a 168 mm (6.61 inches) outside diameter pipe measuring water. The material is carbon 
steel with no liner. These parameters should be entered as follows: 

 

Step 1. Pipe Outside Diameter 

Press 

 or 

, to enter the pipe outside diameter, 

and then press the 

 key.    

 
Note 12:01:00 is the time, Q is the Quality, and R is a diagnostics code.  
Don’t worry about these for now. 
 
 
 
 
 

 
 
 
 