Messages and Diagnostics
SIWAREX FTA
7-145
Error
No.
Data and
command errors -
description
Solution
89
Reserved
90
Reserved
91
Reserved
92
Reserved
93
Reserved
94-
104
Scale command
cannot be
executed
94
Adjustment weight
too small
The adjustment weight in DR 3 is too low. Increase the difference
to the adjustment weight. The measurement value between
sequential adjustment weights is to be at least 5 % of the
measurement range of the load cells apart
95
Reserved
Reserved
96
Tare T exceeded
The subtractive Tare has been exceeded. This error is generated in
the following cases:
With the Tare command, the gross value lies over the defined
permitted tare range.
An externally defined tare value is negative.
An externally defined tare value is greater than the max-value of
weighing range 1 on a multi-segment scale.
An externally defined tare value is greater than the defined max tare
value [%] of the largest weighing range in calibratable operation
(specifications Code OIML in DR 3) of a multi-range scale.
97
Totalising memory
1 deletion not
permitted
The distribution memory cannot be deleted.
The command “Delete distribution memory 1” (114) is not
permitted in calibratable operation. Use command (113) -
“Delete/Log calibratable distribution memory 1”.
The command “Delete calibratable distribution memory 1” (113)
can only be terminated in calibratable operation (Specifications
Code OIML in DR 3) if log text 1 (DR40) contains the distribution
memory 1 field.
Check the parameters.
98
Totalising memory
already deleted
The delete totalizing memory command could not be executed
since the memory is already deleted.
99
Totalising memory
1 could not be
logged
Logging the distribution memory has failed. Check the output
device. Check additional technological faults for indications as to
the cause of the fault. A technology fault is generated indicating
the cause.
100 Weight outside zero
definition range
Zero setting cannot be performed since the current zero point lies
outside the defined zero setting range in DR 3 or the current
gross value lies above the largest defined weighing range value.
101 Scale command
cannot be executed
since not at
standstill 1
The weighing command (Taring, Zero setting, Logging) cannot
be executed because Standstill 1 does not lie within the standstill
time defined in DR 3.
102 Adjustment weight
missing
Adjustment command cannot be executed because the
respective adjustment weight in DR 3 has not been entered.
103 Characteristic curve
shift not possible
Adjustment instruction for characteristic curve shift cannot be
executed. Possible causes: Shift leads to exceeded range,
adjustment not completed, write protect is active).
104
Reserved