SIT SITRANS RD300RANS RD300
Page 44
SITRANS RD300 – OPERATING INSTRUCTIONS
A5E31917845
Low Alarm with Fail-Safe Operation (Set < Reset)
Note: Relay coil is energized in non-alarm condition.
In case of power failure, relay will go to alarm state.
Reset
Set
Input
Relay
LED
Automatic (non-latching)
de-energized
energized
on
off
Relay
LED
Automatic or Manual (non-latching)
ACK
pressed
Relay
LED
Manual (latching)
ACK
pressed
Relay
LED
Manual only after passing above Reset (latching with clear)
ACK
pressed