Installing/Mounting
2.2 F US Clamp-on Communications Module Installation
F US Clamp-on Communications Protocol Manual
Hardware Installation Manual, 11/2015, A5E34981613-AC
11
New Communications Module Installation
This installation procedure is for flowmeters with an I/O Module installed but no Modbus
Communications Module previously installed. Mounting kits are also required for this
installation.
Table 2- 1
Mounting Kit
Mounting Kit
Meter Type
Meter Configuration
1015N-5M-MK3
1010MN
1010N-8M I/O Module with
1010N-2M Module installed
1.
Disconnect power to the unit.
2.
Open the flowmeter cover by releasing the side cover latch
③
.
3.
Loosen the I/O Module cover captive screw
②
and remove the cover, screw and
standoff.
①
Flowmeter
④
Power Connector location
②
I/O Module cover, cover screw,
standoff
⑤
I/O Module
③
Flowmeter Cover Latch
⑥
I/O Module captive mounting screws
Figure 2-1
New Communications Module Installation
4.
Remove the four mounting screws and standoffs
⑥
that secure the I/O Module
⑤
and
discard.
5.
Install the new standoffs supplied in the mounting kit in the vacant positions.