Technical data
Standard cable
(electrode/coil)
Double-shielded cable
(electrode)
Basic data
No. of conductors
3
3
Sqr. area
1.5 mm
2
0.25 mm
2
Screen
Yes
Double
Color code
Brown, blue, black
Brown, blue, black
Outside color
Grey
Grey
Ext. diameter
7.8 mm
8.1 mm
Conductor
Flexible CU
Flexible CU
Isolation material
PVC
PVC
Ambient temperature
Flexible installation
-5 to +70°C
(23 to 158°F)
-5 to +70°C
(23 to 158°F)
Non-flexible installation
-30 to +70°C
(-22 to 158°F)
-30 to +70°C
(-22 to 158°F)
Cable parameter
Capacity
161.50 pF/m
-
Inductance
0.583 μH/m
-
L/R
43.83 µH/Ω
-
For more information on cable lengths, empty pipe detection, and conductivity, see Operating
Instructions for relevant sensor.
11.6
Cable requirements
Coil cable
Electrode cable
Basic data
No. of conductors
2
3
Min. sqr. area
0.5 mm
2
0.2 mm
2
Screen
Yes
Yes
Max. capacitance
N/A
350 pF/m
Max. cable loop resistance
Media temperature:
< 100 °C
40 Ω
N/A
> 200 °C
6 Ω
N/A
Cable glands on sensor
M20x1.5 gland - Cable ø 5 to 13 mm (0.20 to 0.51 inches)
½ NPT gland - cable ø 5 to 9 mm (0.20 to 0.35 inches)
WARNING
Cable glands
For Ex zone 1 installations only certified cable glands with protection type "e" can be used
for the power supply and the coil cable. The cable glands must be approved for the actual
temperature and cable dimension.
Technical data
11.6 Cable requirements
72
SITRANS MAG 6000 I/6000 I Ex de
Operating Instructions, 08/2018,
A5E02083319-AH
Содержание SITRANS F MAG 6000 I
Страница 6: ...Table of contents 6 SITRANS MAG 6000 I 6000 I Ex de Operating Instructions 08 2018 A5E02083319 AH ...
Страница 100: ...Control drawing C 1 Control drawing 100 SITRANS MAG 6000 I 6000 I Ex de Operating Instructions 08 2018 A5E02083319 AH ...
Страница 101: ...Control drawing C 1 Control drawing SITRANS MAG 6000 I 6000 I Ex de Operating Instructions 08 2018 A5E02083319 AH 101 ...
Страница 102: ...Control drawing C 1 Control drawing 102 SITRANS MAG 6000 I 6000 I Ex de Operating Instructions 08 2018 A5E02083319 AH ...