Modbus
Address
Parameter
Default
value
Value
range
Access
level
Description
40151
311
Adjustment Factor
1
- 2 to 2
R/W
Meter correction factor for customer ad‐
justment of the calculated flow value.
Can be used at installations where a
reference meter is used to get reference
to the true flow rate
40196
312
Transducer cable length Product
dependent
0 to 200 m R/W
Cable length (m) from sensor to trans‐
mitter
40173
372
Filter time constant
Product
dependent
0 to 100 s R/W
General time filter for flow measure‐
ment. Filter constant is the time that
must pass before the filter's ouput value
is at 70% of the change of its input val‐
ue. Higher number gives a slower and
more stable flow signal.
40162
373
Low flow cut-off
Product
dependent
0 to 10
R/W
If the flow comes below this percentage
of Qmax, then the flow value is set to
zero
40164
374
Creep lock max number 10
0 to 20
R/W
Alternative cut-off for totalizing and puls‐
ing.
If the totalized flow within this number
(N) of samples exceeds [(N x Qmax x
low flow cut off)/100], the amount is ac‐
cepted for pulsing.
ΣQ > N x Qmax x low flow cut-off/100
● ΣQ: totalized flow within N samples
● N: Creep lock max number
40184
380
Protect linearization
from user access.
No
No, Yes
Read
Protection of the linearization parame‐
ters. It is decided by the order number
and cannot be changed.
The linearization parameters are listed
in the device menu "Linearization"
Totalizer
40673
400
Flow direction totalizer 1 Forward
Forward,
Reverse,
Net
R/W
Calculation principle on flow direction
for forward, reverse or net flow
40675
401
Totalizer 1 change date Last
change
date
dd-mm-
yyyy;
hh:mm:ss
Read
Date and time when totalizer 1 function
was changed
40674
410
Flow direction totalizer 2 Reverse
Forward,
Reverse,
Net
R/W
Calculation principle on flow direction
for forward - reverse or net flow
40678
411
Totalizer 2 change date Last
change
date
dd-mm-
yyyy;
hh:mm:ss
Read
Date and time when totalizer 2 function
was changed
Pipe data
40181
300
Number of paths
Product
dependent
2
R/W
Number of paths on the sensor
40149
302
Max sample frequency 15
15
Read
The maximum frequency for flow meas‐
urement
Parameter lists
A.4 Meter setup
FUS380/FUE380
94
Operating Instructions, 09/2016, A5E00730100-AB
Содержание SITRANS F FUE380
Страница 6: ...Table of contents FUS380 FUE380 6 Operating Instructions 09 2016 A5E00730100 AB ...
Страница 20: ...Description 3 5 Principle of operation FUS380 FUE380 20 Operating Instructions 09 2016 A5E00730100 AB ...
Страница 62: ...Functions 7 4 Hardware key FUS380 FUE380 62 Operating Instructions 09 2016 A5E00730100 AB ...
Страница 78: ...Troubleshooting FAQs 9 2 Diagnosing with PDM FUS380 FUE380 78 Operating Instructions 09 2016 A5E00730100 AB ...
Страница 100: ...Settings B 3 Ordering of spare parts FUS380 FUE380 100 Operating Instructions 09 2016 A5E00730100 AB ...