①
Gain step
②
Gain step max
③
Gain step min
④
Transducer distance [mm]
Figure 9-2
SITRANS FUS080 Gain step vs. Transducer distance
Gain step
The graph above shows which gain step to expect for a particular transducer distance (face
to face distance). The upper curve indicates the upper limit and the lower curve indicates the
lower limit.
The graph applies to perfectly aligned transducers in clean water, but takes into account the
expected variance from the converter and the transducers.
Gain step error conditions
Condition
Cause
15
● No transducer (e.g. defective cable)
● Empty pipe
High values
● High damping (e.g. inhomogeneous media)
● Unsuitable medium
● Path angular misalignment
Unstable/fluctuating values
● Air bubbles or solids in medium
● Bad inlet conditions
Δ gain step between path 1 and path 2 > 1
(2-path systems only)
● Bad inlet conditions
● Inhomogeneous medium
● Path angular misalignment
Troubleshooting/FAQs
9.2 Diagnosing with PDM
FUS380/FUE380
76
Operating Instructions, 09/2016, A5E00730100-AB
Содержание SITRANS F FUE380
Страница 6: ...Table of contents FUS380 FUE380 6 Operating Instructions 09 2016 A5E00730100 AB ...
Страница 20: ...Description 3 5 Principle of operation FUS380 FUE380 20 Operating Instructions 09 2016 A5E00730100 AB ...
Страница 62: ...Functions 7 4 Hardware key FUS380 FUE380 62 Operating Instructions 09 2016 A5E00730100 AB ...
Страница 78: ...Troubleshooting FAQs 9 2 Diagnosing with PDM FUS380 FUE380 78 Operating Instructions 09 2016 A5E00730100 AB ...
Страница 100: ...Settings B 3 Ordering of spare parts FUS380 FUE380 100 Operating Instructions 09 2016 A5E00730100 AB ...