Precautions
Avoid installation at the highest point in the system because air bubbles will be trapped in the
flowmeter.
Avoid installation at a point where there is a free outlet after the flowmeter.
The flowmeter pipe section may be installed in either a horizontal or vertical position.
Note
To obtain maximum battery lifetime of the Lithium Thionyl Chloride battery pack, Siemens
recommends installing the flowmeter transmitter in an upright position.
Installing/Mounting
4.2 Sensor installation
FUS380/FUE380
Operating Instructions, 09/2016, A5E00730100-AB
25