To set segment
6 (
°
C)
Temperature from 55
°
C to 30
°
C
Duration 2h - slope - gradient 0.21
°
C/min
hh : mm : ss
Durat.
2 : 00 : 0
Wait duration
: ON
G1
Channel 1
Control
: ON
Wait set point
: ON
Max speed
: OFF
Set point
: 30.0 0
°
C
Grad.
: 0.21
°
C/min
Near set
: 1,00
°
C
G2
G3
Because the same temperature value
is maintained
OP2605-090
Channel 2
Control
: ON
Wait set point
: OFF
Max speed
: ON
G4
G5
OP2605-100
Set point
: 95.00 RH %
Grad.
: 20,0 RH %/min
Near set
: 3,00 RH %
G8
to save the programmed segment
Save segment
: SAVE
G9
EDIT PROGRAM
< ESC > Prev. menu
Segment N. 7
Wait for this number to appear
Water recharge:
OFF
Not used:
OFF
Not used:
OFF
SP. hum. CTRL.
ON
To improve the system performance
(experimentally set)
G6
To set segment
6 (% RH)
Humidity from 90% to 95% and maintenance
Duration 2h - Maintenance
30
55
6
Unreal gradient which enables maxi-
mum speed to be obtained
Aux 1:
OFF
Aux 2:
OFF
Aux 3:
OFF
Aux 4:
OFF
In this segment the AUX1 must be
switched OFF
G8
SHIFT
ENTER
• Press
repeatedly
WARNING!
If you have accidentally saved the same segment twice, correct the number on this page.
This page is visualized.
Save segment
: OFF
Total segment
: 0
Confirm progr
: OFF
When this page is visualized, if the previ-
ously described operations have been
carried out, the segment has been saved
and therefore must not be saved again.
Move on to the next step.
90
95
6