21
Turn all the values to ON as shown
SHIFT
B8
Channel 1
Control
: ON
Wait set point
: ON
Max speed
: ON
ENTER
B9
Set point
: 0.00
°
C
Grad.
: 0.00
°
C/min
Near set
: 1,00
°
C
B10
Set point
: 30.00
°
C
Grad.
: 0.00
°
C/min
Near set
: 1,00
°
C
To set segment
1 (
°
C)
Temperature 30
°
C
At maximum speed
Leave the gradient at 0 = max speed
Default value (tollerance within which
the segment is considered to have been
reached)
Channel 2
Control
: OFF
Wait set point
: OFF
Max speed
: OFF
Channel 2: to set the humidity
Turn all the values to ON as shown
SHIFT
B12
Channel 2
Control
: ON
Wait set point
: ON
Max speed
: ON
ENTER
20
40
30
1
1,5 2,5
2
1
OP2605-050
Set point
: 0.00 RH %
Grad.
: 0.00 RH %/min
Near set
: 3,00 RH %
1
B13
B14
Set point
: 60.00 RH %
Grad.
: 0.00 RH %/min
Near set
: 3,00 RH %
To set segment
1 (% RH)
60% humidity
At maximum speed
Leave the gradient at 0 = max speed
Default value
(tollerance within which the segment is
considered to have been reached)
40
60
80
90
OP2605-060
1
• Enter the values shown on
the page.
B11
Default value