3.2
STEP C
Set a controlled humidity slope from 60% to 90% with a gradient of 2%/min with a
constant temperature and 60% RH humidity
• Wait for the chamber to reach the preset values, 35
°
C
and 60% RH (Step A and Step B).
• Follow the example given to programme a temperature
slope as shown in the diagram with the gradient indi-
cated.
• Move to the next page.
Temperature slope
already carried out
K2
Settings
A1
Temp. Set
35.0
°
C
Humid. Set
60.0 %
A2
The previously set values are visual-
ized.
Temp. gr.
0.00
°
C min
Hum. gr.
2.00 RH%min
• Turn the temperature gra-
dient to 0
• Turn the humidity gradi-
ent to 2
A4
ENTER
A3
Temp. Set
35.0
°
C
Humid. Set
90.0 %
ENTER
If the K1 key lights up, the machine will
carry out the command immediately
Final humidity
40
50
60
70
80
90
30
15’
20’
25’
t= min.
% RH
KEYS
DISPLAY
FUNCTIONS
gradient:
2%/min
30’