Theory of Operation
February 5, 2004
2-9
The DirectRay
Detector
The DirectRay Detector, illustrated in Figure 2-5, directly converts the X-rays to
digital data. The DirectRay Detector has several layers. The first layer inside the
cover, the SE-TFT detector matrix, converts the X-rays to digital data.
The DirectRay Detector also contains the control electronics and the precision
analog electronics that perform the data capture and conversion into digital
format. An 8-bit micro-controller in the DirectRay Detector performs the
management functions of setup, monitoring, and control of the acquisition
electronics.
Figure 2-5.
DirectRay Detector Structure
INTERNAL
DETECTOR
COVER
X-RAY
CONVERSION
LAYER
DETECTOR (TFT)
MATRIX
SUPPORT
STRUCTURE
READOUT
CHARGE
AMPLIFIERS
READOUT
GATE
DRIVERS
FRAME
IMAGE AREA
CARBON FIBER
COMPOSITE
COVER
ELECTRONICS HOUSING, CONTAINS:
A/D CONVERTORS
MULTIPLEXER
HIGH VOLTAGE POWER SUPPLY
DC-DC CONVERTER
ANTI-SCATTER
GRID MECHANISM
Содержание MAMMOMAT NovationDR
Страница 7: ...vi February 5 2004 FFDM S Service Manual THIS PAGE IS INTENTIONALLY BLANK ...
Страница 29: ...2 12 February 5 2004 FFDM S Service Manual This Page Left Intentionally Blank ...
Страница 47: ...4 10 February 5 2004 FFDM S Service Manual ...
Страница 51: ...4 14 February 5 2004 FFDM S Service Manual This Page Left Intentionally Blank ...
Страница 68: ...PXCM Web Tool February 5 2004 5 17 ...
Страница 76: ...PXCM Web Tool February 5 2004 5 25 ...
Страница 83: ...6 4 February 5 2004 FFDM S Service Manual This Page Left Intentionally Blank ...
Страница 95: ...B 2 February 5 2004 FFDM S Service Manual This Page Left Intentionally Blank ...