Siemens AG
SPB7-250.812.03.02.02
MAMMOMAT Novation
08.04
CS SD 24
Installing the system components
29
Page 29 of 36
Medical Solutions
NOTE
The base plates shall be secured with at least four screws each
(two on either side).
NOTE
Max pull forces that occur per mounting point: 1.5
kN
Fig. 21: Base plate for the radiation shield
Pos. 1
screws
Pos. 2
Room for calble lead-in in floor
1. Position the stand/radiation shield and mark out appropriate fixing holes on the floor us-
ing the base plate as a template.
2. Drill the holes and insert dowels and screws.Tighten the screws to 10 Nm.
Example of the leveling screws
0
NOTE
If safety screws are used, the manufacturer’s specifications and
instructions must be observed
With installation on the solid floor, cut out the floor covering. Con-
crete quality for the solid floor, min. C20/25.
Mounting materials must be obtained locally