Installing the system components
21
Page 21 of 36
Medical Solutions
5-
4Installing the system components
Room planning
0
NOTE
•
min. 150 mm if the unit is not secured to the floor (in a special
service situation, the generator must be swung out from be-
hind at the stand; for this, the stand must be moved 500 mm
away from the wall).
min. 500 mm if the unit is secured to the floor.
•
If needed, additional cable conduits, Part No. 64 38 795 X041E,
can be ordered for the connection between the stand and the
radiation shield.
•
For safety reasons, the acquisition workstation must be set up
outside the area, per IEC 60601-1-1
Fig. 12: Room planing