Siemens AG
SPB7-250.812.03.02.02
MAMMOMAT Novation
08.04
CS SD 24
Unpacking the system components
17
Page 17 of 36
Medical Solutions
Rotary motion
0
Fig. 8:
Transport safeguard for rotary motion
Pos. 1
screw
1. Remove the transport safeguard screw
with spacer to enable rotary
motion. The screw and the spacer should be kept for the mounting of the lifting carriage
transport safeguard.
2. Remove the transport safeguard on the balancing spring.