IONPURE
®
VNX50HH-2 CEDI Modules
Page 24
IP-MAN-VNX50HH2
2.
Turn on the feed water supply to the system. Slowly open the feed valve to allow
RO product water to flow through the module to drain.
3.
After flushing all residual cleaning solutions from the system, follow the initial
start-up procedures in Section 5 of this manual. Start without the DC power
applied to the VNX module for five minutes and then restore normal DC power.
6.9. Cleaning with 5% Brine + 1% Caustic Solution
Sodium Hydroxide is a corrosive chemical. Wear glasses, gloves, and
protective clothing when using this or any corrosive chemicals.
A 5% sodium chloride (NaCl)/1% sodium hydroxide (NaOH) solution works well for
removing organic foulants from the module.
Also see Ionpure.com for Service
Bulletin IPSB-2007-2.
This procedure involves two steps:
1. Brine/caustic
cleaning.
2. Water
flush.
STEP 1: Cleaning Solution Recirculation
This step removes organic foulants from the module.
Chemicals necessary for Step 1:
VNX50-HH
Water (RO permeate or CEDI product)
99 gal (375 liters)
Sodium chloride (NaCl)
42 lbs (18.75 kg)
Sodium hydroxide (NaOH) pellets
OR 50% NaOH
8.25 lbs (375 kg)
1.29 gal (4.8 liters)
NOTE:
The following cleaning procedures are based on the piping setup shown in
1. Prepare the VNX system for chemical cleaning. Follow the instructions in Section
6.5. Make sure the cleaning tank drain valve is closed.
2. Fill the tank with the required amount of water.
3. Open the cleaning pump bypass valve and suction valve completely.
4. Start the cleaning pump.
5. Dissolve the specified amount of sodium chloride (NaCl) in the water.
6. Dissolve or mix the specified amount of sodium hydroxide (NaOH) in the water.
7. When chemicals are well mixed, gradually open the cleaning pump discharge valve
while closing the bypass valve to adjust the product and reject flow rates to the
values given in Section 6.5.
8. Recirculate the solution through the module for 30-60 minutes.
9. Turn off the cleaning pump.
10. Check pH. Neutralize if necessary. Drain the tank.
Содержание IONPURE VNX50HH-2
Страница 40: ...IONPURE VNX50HH 2 CEDI Modules Page 40 IP MAN VNX50HH2 APPENDIX B VNX50HH LAYOUT ELEVATION DRAWING B 1 ...
Страница 41: ...IONPURE VNX50HH 2 CEDI Modules Page 41 IP MAN VNX50HH2 APPENDIX B VNX50 HH PID DRAWING B 2 ...
Страница 42: ...IONPURE VNX50HH 2 CEDI Modules Page 42 IP MAN VNX50HH2 APPENDIX B VNX50 HH ELECTRICAL CONNECTIONS DRAWING B 3 ...
Страница 46: ...IONPURE VNX50HH 2 CEDI Modules Page 46 IP MAN VNX50HH2 APPENDIX C VNX50 HH PP PLUG DRAWING C 4 ...