Appendi x A Technical Des cription
A - 20
Instructions for Use
Contrast Agent Imaging Option
WARNING:
Carefully follow the manufacturers’ instructions for use, including indications and
contraindications, when administering ultrasound contrast agents.
Compatible transducer: 4C1
Supported studies: Abdomen, Renal, Gynecology, Pelvis
Compatible transducer: 9L4
Supported studies: All except Fetal Echo, Obstetrics, Pediatric Abdomen, and Pediatric Hip
Compatible transducer: EC9-4, MC9-4
Supported studies: Prostate
Compatible transducer: 6C1 HD
Supported studies: Abdomen, Renal
Supports the following acquisition techniques: Contrast Pulse Sequencing (CPS), Agent
Detection Imaging (ADI)
–
CPS is a low Mechanical Index continuous real-time, non-linear imaging technique
that provides contrast-to-tissue specificity.
–
ADI is a high Mechanical Index imaging technique that uses the emission properties of
contrast agents to characterize their presence or absence in the tissue.
Supports the following contrast agent destruction techniques: Burst, Micro-Bubble
Destruction (MBD)
–
Burst is a high Mechanical Index technique designed to rapidly destroy the injected
contrast agent bubbles for a view of re-perfusion immediately afterwards.
–
MBD is a high Mechanical Index non-imaging mode technique for rapid destruction of
contrast agent.
Содержание Acuson S2000
Страница 12: ...1 Introduction 1 2 Instructions for Use ...
Страница 14: ...1 Introduction 1 4 Instructions for Use System Review Example of the ultrasound system ...
Страница 84: ...2 Safety and Care 2 54 Instructions for Use ...
Страница 86: ...3 System Setup 3 2 Instructions for Use ...
Страница 112: ...3 System Setup 3 28 Instructions for Use ...
Страница 114: ...4 Examination Fundamentals 4 2 Instructions for Use ...
Страница 144: ...5 Transducer Accessories and Biopsy 5 2 Instructions for Use ...
Страница 196: ...7 Specialty Transducers 7 2 Instructions for Use ...
Страница 200: ...7 Specialty Transducers 7 6 Instructions for Use ...
Страница 202: ...8 Physiologic Function 8 2 Instructions for Use ...
Страница 208: ...9 eSieFusion Imaging 9 2 Instructions for Use ...
Страница 236: ...10 Virtual Touch Applications 10 2 Instructions for Use ...
Страница 258: ...10 Virtual Touch Applications 10 24 Instructions for Use ...
Страница 302: ...Appendix A Technical Description A 44 Instructions for Use ...
Страница 326: ...Appendix B Control Panel and Touch Screen B 24 Instructions for Use ...
Страница 328: ...Appendix C Control Panel C 2 Instructions for Use ...
Страница 394: ...Appendix D On screen Controls D 50 Instructions for Use ...
Страница 444: ...Appendix F Acoustic Output Reference F 2 Instructions for Use ...
Страница 516: ...Appendix F Acoustic Output Reference F 74 Instructions for Use ...
Страница 517: ......
Страница 518: ......