ULTRAMAT 6 / OXYMAT 6 / CALOMAT 6 / ULTRAMAT 23 / OXYMAT 61
Schnittstelle PROFIBUS DP / PA Interface
A5E00054148-03
92
On ULTRAMAT 23 there are the following possibilities due to the O
2
measuring components (sensor):
Variants
IR measuring
component 1
IR measuring
component 2
IR measuring
component 3
O
2
(if applicable)
7MB 2331
TB/AI 1
----------------------
----------------------
TB/AI 2
7MB 2333
TB/AI 1
TB/AI 2
----------------------
TB/AI 3
7MB 2334
TB/AI 1
TB/AI 2
TB/AI 3
TB/AI 4
The corresponding gas for the measuring components is indicated by the device MLFB number (see rating
plate).
3.1.1
Transducer Block with ULTRAMAT, OXYMAT and CALOMAT
The analyzer transducer block receives its input value from the ULTRAMAT, OXYMAT and CALOMAT basic
unit. This causes the device-related functions of sensors as well as the measured-value acquisition and
processing to be separate from the function block.
Rel.
Index
Parameter
Description
8
COMPONENT_NAME
Name of the component
9
PV
Measured value which is forwarded to the AI block
10
PV_UNIT
Measured value unit
11
PV_UNIT_TEXT
Unit in text form
12
ACTIVE RANGE
Current measuring range/always "1". ULTRAMAT, OXYMAT
and CALOMAT use the largest measuring range here
(characteristic curve upper value)
13
AUTORANGE_ON
Automatic measuring range switchover TRUE/FALSE
No significance as ULTRAMAT, OXYMAT and CALOMAT
always use the largest measuring range (characteristic curve
upper value).
14
SAMPLING_RATE
Time between two measured value readings
25
NUMBER_OF_RANGES
Number of measuring ranges/always just one measuring range
26
RANGE_1
Measuring range lower and upper value (“largest measuring
range“)
34
POSITIONING_VALUE
Value (1 byte) for triggering autocalibration. Taken from DO
block (OUT_D).
35
FEEDBACK
Value (1 byte) for feedback on the autocalibration status (taken
from DO block FEEDBACK).