ULTRAMAT 6 / OXYMAT 6 / CALOMAT 6 / ULTRAMAT 23 / OXYMAT 61
Schnittstelle PROFIBUS DP / PA Interface
A5E00054148-03
88
2.8.2
Installing the Optional ULTRAMAT / OXYMAT / CALOMAT 6 Field Housing Module
☛
Note
The
A5E00015899 Exi
module for intrinsically safe connection to the PROFIBUS-PA may only be
installed in
Exp[ia/ib]
field housing devices.
With all other non-Ex devices only the
A5E00034504
for non-intrinsically safe connection to
PROFIBUS-PA or the
A5E00019145
module for connection to the PROFIBUS-DP can be installed.
Installation instructions:
•
Disconnect the device from the mains
•
Open the left-hand housing door
•
Remove the sheet metal cover
•
Remove the sheet metal cassette from the device
•
Remove the old optional module if installed
•
Fit the module in the rail
•
Fasten the module with the M3 screws to the underside of the cassette between connectors
•
Fit the bracket on the top edge of the optional module
•
Connect the connectors to the motherboard
•
Fit the sheet metal cassette in the device
•
Connect the 37 Sub-D-connector with the terminal block
•
Replace sheet metal cover
•
Connect the device to the PROFIBUS
•
Connect the device to the mains
Cassette
Motherboard
Connect 26-pin
ribbon cable
Optional module
Fig. 2.2