
ULTRAMAT 6 / OXYMAT 6 / CALOMAT 6 / ULTRAMAT 23 / OXYMAT 61
Schnittstelle PROFIBUS DP / PA Interface
A5E00054148-03
119
7.4.2
CALOMAT 6 Diagnostic Information
Fault (S)
Maintenance
request (W)
No.
Meaning of fault/maintenance request Diagnostic
byte/bit
Diagnosis bit name
S1
Parameter memory
1 / 4
2 / 2
DIA_MEM_CHKSUM
DIA_CONF_INVAL
S4
External fault
1 / 1
DIA_HW_MECH
S5
TCD temperature
1 / 1
DIA_HW_MECH
S12
Power supply
2 / 1
DIA_SUPPLY
S14
Measured value > characteristic curve
upper value
1 / 1
1 / 5
DIA_HW_MECH
DIA_MEASUREMENT
S15
Calibration cancelled
1 / 5
DIA_MEASUREMENT
W1
Calibration difference too large
2 / 5
DIA_MAINTENANCE
W4
Set clock
2 / 5
DIA_MAINTENANCE
W6
Temperature sensor LCD display
1 / 3
2 / 5
DIA_TEMP_ELECTR
DIA_MAINTENANCE
W9
External maintenance request
2 / 5
DIA_MAINTENANCE
W10
Autocal check difference
2 / 5
DIA_MAINTENANCE