34
16.2 Electrical Data
Voltage in:
230V±10%, ~50/60Hz.
Current
in:
1.25A
Mains cable fuse:
3-5A
Duty cycle:*
10%
2 mins
of continuous use followed by
18 mins
not in use.
*Electrically operated beds are intended to be operated intermittently rather than continuously. If the bed is operated
continuously for up to 2 minutes it must then be left for at least 18 minutes before reuse to allow the electrical system to cool
sufficiently.
If the bed is continuously used for an extended period of time and it exceeds the duty cycle the control
box may become temporarily disabled or irreparably damaged.
Safety standards:
IEC 60601-1: 2005
IEC 60601-1-2:2007
IEC 60601-2-52:2009
The Solite UK Bed is compliant to the requirements of 60601-1-2: 2007, the RF
emissions are very low and are not likely to cause any interference to nearby electronic
equipment, however interference to sensitive equipment is possible.
Qualified users of the Solite UK Bed can help prevent electromagnetic interference
by maintaining a suitable distance between portable and mobile RF communications
equipment (transmitters). If the Solite UK Bed or any alternative equipment is found to
be operating abnormally mitigation measures are to be taken, such as the separation
distances being increased and / or the products being reorientated.
Electrical
shock
prevention:
Applied parts:
Mattress platform
Profiling
sections
Upper half of head / foot ends
Handset
Liquid ingress protection:
IPX4 - Splash resistant
Battery backup:
Individual batteries non replaceable
(SOLITE/UK/4/BB only)
Noise level:
59dB(A) max
Service life:
5 years
Environmental conditions:
Operational Limits
Transportation / Storage Limits
Ambient Temperature
+5ºC to +40°C
-10°C to +50ºC
Humidity
20% - 90% at 30°C
non condensing
20% - 90% at 30°C
non condensing
Atmospheric Pressure
800 to 1060hPa
800 to 1060hPa
Altitude
≤ 2000m
≤2000m