16. SPECIFICATION

16.1 Bed Data

Overall length:   

 

 

          2250mm - 2395mm

Overall width: 

 

 

 

          970mm

Mattress platform height:   

 

          300mm – 705mm*

Under bed clearance (to underside of frame):         250mm
Under bed clearance (to underside of actuator):   105mm
Mattress platform length:   

 

          2000mm - 2145mm

Mattress platform width:   

 

          927mm

Head down tilt: ** 

 

 

           0 - 11°

Foot down tilt:   

 

 

           0 - 11°

Mattress platform angles (max):

Safe working load: 

 

 

           222kgs (35 stone)

Maximum patient weight: 

 

           191kgs (30 stone)

Product 

weight: 

   

 

(On transport stand):  

 

           75kgs

Individual item weight:
 

Backrest : 

 

 

           19.5kgs

 

Leg section:  

 

 

           19kgs

 

Bed end: 

 

 

           17kgs

 

 

 

Application environment:   

 

            3 and 4

Shock and vibration: 

 

 

            To be used on a flat level floor   

 

 

 

 

                                (e.g. vinyl/carpet/laminate based)

UV: 

 

 

 

 

             Intended for indoor use only

70°

27°

24°

Warning

•  *If patient requirements / situation are such that a higher platform  

 

 height 

is required please contact your provider or Sidhil Ltd. (see  

 

  section 2) to discuss suitability of alternative bed frames.
•  **If patient requirements are such that Trendelenburg functionality    
  is deemed to pose a potential risk, a replacement handset can  

 

  be purchased with the Trendelenburg function removed (handset 
  code CONTROL/127), please refer to the contact information  

 

  in section 2 to order or to request further information – Sidhil  

 

  recommend the use of this handset when the bed is being used in a  

 

  domestic environment.

Содержание Solite UK

Страница 1: ...Instructions for use Community Care Bed Range www sidhil com...

Страница 2: ...levels of environmental awareness and sustainability across our manufacturing operation At Sidhil everything we do is designed around quality From our modern and efficient manufacturing plant in Halif...

Страница 3: ...ecking the Bed 16 9 OPERATION OF THE BED 17 9 1 Operational Limits 17 9 2 Installation Preparing For Use 17 9 3 Brake System 19 9 4 Side Rails and Mattresses 20 9 5 Side Rail Safety 20 9 6 Electrical...

Страница 4: ...contact your provider or if in doubt contact Sidhil Ltd at the following address Please quote the Solite UK serial number on all correspondence Sidhil Ltd Sidhil Business Park Holmfield Halifax West Y...

Страница 5: ...designed to provide patients with optimum independence and freedom of movement through the use of a touch button handset Thebedoffersgreatlyreducedmanualhandlingrequirementsforthecarerbyproviding a b...

Страница 6: ...to four separate sections Transport stand to aid storage and bed transportation Electrical system IPX4 rated Splash resistant Battery backup functionality SOLITE UK 4 BB only If patient requirements a...

Страница 7: ...Before a patient uses the bed a risk assessment must be performed on a patient by patient basis The risk assessment should include but is not limited to Entrapment Falling out of the bed Children and...

Страница 8: ...xplosion fire Bed functions must be locked out if there is any doubt about the ability of the patient to operate the bed safely If children adults with learning difficulties or even pets pose a potent...

Страница 9: ...on their side Ambient temperature 10 C to 50 C Humidity 20 90 at 30 C not condensing Atmospheric pressure 800hPa to 1060hPa altitude 2000m The bed is not intended for patient transport it is not to be...

Страница 10: ...mage is noticed to any electrical component turn off at the mains and contact your provider or Sidhil Ltd immediately Mattress suitability Refer to section 17 Type B Applied Part Applied Part The part...

Страница 11: ...ket 3 Castor 4 Leg section 5 Control box back rest actuator 6 Leg rest actuator 7 Extension hand wheels 8 Locking collar 9 High low actuator 10 Backrest section Transport stand not shown 7 PARTS IDENT...

Страница 12: ...carefully place it flat on the floor Loosen the hand wheels on the leg section attaching it to the transport stand and then lift this section off the stand and carefully place it flat on the floor Tur...

Страница 13: ...rs Bring both halves of the bed together and align each section so that the spigots in the backrest section locate into the open tube ends in the leg section Pull the two sections together and tighten...

Страница 14: ...able has enough flex to allow full movement of the actuator Note The two bed ends are identical however plugging them into the correct port is important Note the location of the correct end before plu...

Страница 15: ...unction is not functioning as expected the cables may have been incorrectly located double check against the diagram on the control box Caution 8 5 Ancillary Product Cable Routing When fitting ancilla...

Страница 16: ...slack in the cables to allow for movement Is the bed clear of obstructions Have the 2 clevis pins been passed through the bed s central joint and secured with R clips Has a risk assessment been perfo...

Страница 17: ...mbs are clear of moving parts The mains plug is the disconnect device for the means of isolating the bed from the mains supply the plug must be accessible at all times Ensure the mains cable is plugge...

Страница 18: ...tc being evident to minimise the risk of them being swallowed by a baby child or the bed occupant and creating a choking hazard Keep the bed away from sources of heat and naked flames e g cigarettes f...

Страница 19: ...obstacles such as over bed tables and other furniture are not causing an obstruction Ensure the bed is positioned an appropriate distance from walls other furniture to prevent damage or patient injur...

Страница 20: ...fe Side side rail in some circumstances It is the equipment provider s responsibility to ensure suitability for use Whilst every care has been taken to ensure that the design of Sidhil s side rails me...

Страница 21: ...l Warning Caution Ensure a risk assessment is undertaken to ensure the suitability of the occupant using the handset If patient requirementsare such that Trendelenburgfunctionality is deemed to pose a...

Страница 22: ...function if a patient could be injured due to inadvertent motion of the mattress platform The lockout function on the handset is to be used at the discretion of the carer If children adults with lear...

Страница 23: ...ght not important Lift the leg section a little so that the support bow releases from the channel and rests on the supporting tube Gently lower the leg section down The knee break angle will now becom...

Страница 24: ...its clip To Extend the Platform Loosen by 3 4 turns the two hand wheels located on the platform section Slide the platform extension out to the correct length to suit the mattress and extension block...

Страница 25: ...g both halves of the bed frame split the bed in half and gently lower onto the floor Unlock both locking collars on one end collar must be lifted before turning Whilst supporting the bed end lift the...

Страница 26: ...through the larger open tubes on the side of the transport stand brackets ensuring the electrics are facing inwards Carefully lift the leg section and lower the open end onto the vertical tubes ensuri...

Страница 27: ...on Hold support the section Remove the pins that hold the actuator in place Gently lower the section s to the flattened position 12 BATTERY CHARACTERISTICS SOLITE UK 4 BB ONLY The internal battery tri...

Страница 28: ...atically moving to the dirtiest parts Extra care should be taken around areas where excess dirt or dust may gather The cloth should be changed during the cleaning process if it becomes soiled Rinse do...

Страница 29: ...team cleaned The individual manufacturer s instructions shouldbefollowedwhenusingasteamcleanerandthefollowingprecautionsobserved Avoid directing steam directly at electrical components and reduce stea...

Страница 30: ...on light on the control box is lit if not replace fuse Actuator handset leads not plugged in Check plug connections on the control box and actuators Damage to mains cable actuator cable or handset cab...

Страница 31: ...n the handset Check that all electrical cables are in good condition Check the retaining clip is fastened to the control box securing the electrical cables in place Checkthattheredretainingringiscorre...

Страница 32: ...e bed frame is not to be disposed of in general municipal waste Some of the electrical components could be harmful to the environment and where viable the components can be recovered and reused recycl...

Страница 33: ...ication environment 3 and 4 Shock and vibration To be used on a flat level floor e g vinyl carpet laminate based UV Intended for indoor use only 70 27 24 Warning If patient requirements situation are...

Страница 34: ...ronic equipment however interference to sensitive equipment is possible Qualified users of the Solite UK Bed can help prevent electromagnetic interference by maintaining a suitable distance between po...

Страница 35: ...side side rails SOLITE SR 2 Grange side rails GRANGE SR Universal standard side rails UNIVERSAL SR MAG Universal high side rails UNIVERSAL SR HI MAG Side rail pads 1316 PS CREAM Universal high side r...

Страница 36: ...mm 152mm 53 58kg m Softrest Contour MAT SOFT CON 1990mm 864mm 152mm 35 37kg m Softrest Thin MAT SOFT THIN 1990mm 864mm 64mm 38 40kg m Essential Foam MAT BASIC 1990mm 864mm 127mm 30 33kg m Foam Extensi...

Страница 37: ...the top of the side rail It is essential that a patient risk assessment is performed to ensure that this does not introduce a hazard to the patient Sidhil recommends setting the pressure of the Solo I...

Страница 38: ...provided by law a correct the defect by product repair without charge for parts and labour b replace the product with one of the same or similar design or c refund the purchase price All replaced part...

Страница 39: ...39 NOTES...

Страница 40: ...STRUC SOLITE UK 16 01 2015 Rev 9 CONTACT INFORMATION Tel 44 0 01422 233000 Fax 44 0 01422 233010 Email sales sidhil com www sidhil com Sidhil Business Park Holmfield Halifax HX2 9TN A member of the Si...