FOR YOUR SAFETY
2.1.2
Dangerous substances
Hazard through acidic liquids
If the gas sensor is mechanically damaged, corrosive liquid may escape and cause serious
injury.
●
Take appropriate protective measures for work (e.g., by wearing a safety mask,
protective gloves and acid resistant clothes).
●
In case of contact with the skin or eyes, rinse the affected parts immediately with clear
water and consult a doctor.
Hazard through escaping calibration gases
Incorrect handling of calibration gases can cause high concentrations above the workplace
exposure limits to escape from the gas sensor into the ambient air and lead to poisoning.
●
Observe the Safety Datasheets of the calibration gases.
●
Ensure adequate ventilation.
●
Limit gassing time.
2.2
Intended use
The Gas Adjustment Kit serves to check and adjust the gas sensors in the VISIC100SF.
2.3
Requirements for personnel qualification
The device may only be operated by authorized persons who, based on their training on,
and knowledge of the specific device, as well as knowledge of the relevant regulations can
assess the tasks given and recognize the hazards involved.